DMM commands only work when the device is switched to/through the DMM mode manually first.
:DMM:MEAS? - get the current measurement readout
:DMM:CONFIGURE:VOLTAGE DC/AC → switch to voltage measurement, switch to DC or AC mode
:DMM:CONFIGURE:VOLTAGE? returns it out again, or error if not in voltage mode
:DMM:CONFIGURE:CURRENT DC/AC → switch to current measurement, switch to DC or AC mode
:DMM:CONFIGURE:CURRENT? returns it out again, or error if not in current mode
:DMM <val>
- val:
DIODe, CAPacitance, RESistance, CONTinuity
:DMM? → returns mode: R (resistance)/RS (continuity)/DIODE (diode)/C (capacitance)/error (if voltage or current)
:DMM:RANGE? → range readout
:DMM:RANGE ON → cycle through ranges (for non-voltage/current)
:DMM:RANGE V/MV/A/MA → set the respective range (for voltage/current)
:DMM:AUTO? → range readout
:DMM:AUTO ON/OFF → toggle automatic ranging on/off
DMM commands only work when the device is switched to/through the DMM mode manually first.
:DMM:MEAS?- get the current measurement readout:DMM:CONFIGURE:VOLTAGE DC/AC→ switch to voltage measurement, switch to DC or AC mode:DMM:CONFIGURE:VOLTAGE?returns it out again, orerrorif not in voltage mode:DMM:CONFIGURE:CURRENT DC/AC→ switch to current measurement, switch to DC or AC mode:DMM:CONFIGURE:CURRENT?returns it out again, orerrorif not in current mode:DMM <val>DIODe,CAPacitance,RESistance,CONTinuity:DMM?→ returns mode: R (resistance)/RS (continuity)/DIODE (diode)/C (capacitance)/error (if voltage or current):DMM:RANGE?→ range readout:DMM:RANGE ON→ cycle through ranges (for non-voltage/current):DMM:RANGE V/MV/A/MA→ set the respective range (for voltage/current):DMM:AUTO?→ range readout:DMM:AUTO ON/OFF→ toggle automatic ranging on/off